Loading...
Resolution Limits of Time-of-Flight Mass Spectrometry with Pulsed Source
Qu, Guangzhi
Qu, Guangzhi
Abstract
Matrix Assisted Laser Desorption/Ionization (MALDI) is a time-of-flight mass spectrometry commonly used to detect a wide mass range of biomarkers. However, MALDI requires a high laser pulse energy to create ions with a mass higher than 50,000 Daltons. That high laser energy increases the net ion production but it also degrades the instrument's mass resolution. This project uses a Room Temperature Ionization Liquid (RTIL) as a liquid matrix with a self healing surface instead of a standard crystal matrix to increase shot to shot reproducibility, enabling a systematic study of the origin of the resolution degradation. This study shows that the main source of the resolution degradation is the ionic space charge which delays the ejection of ions into the acceleration region, essentially increasing the ionization pulse time to be as long as hundreds of nanoseconds. This study includes simulation and experimental results to document this effect.
Description
Date
2016-04-01
Journal Title
Journal ISSN
Volume Title
Publisher
Collections
Download Dataset
Files
Loading...
Qu_wmgrad_0261E_10073.pdf
Adobe PDF, 3.33 MB
Rights Holder
Usage License
Embargo
Research Projects
Organizational Units
Journal Issue
Keywords
Citation
Department
Physics
DOI
http://doi.org/10.21220/S2F304